Intra-session and inter-day reliability of the Myon 320 electromyography system during sub-maximal contractions

Graeme G. Sorbie, Michael J. Williams, David W. Boyle, Alexander Gray, James Brouner, Neil Gibson, Julien S. Baker, Chris Easton, Ukadike C. Ugbolue

Research output: Chapter in Book/Report/Conference proceedingChapter

Abstract

Electromyography systems are widely used within the field of scientific and clinical practices. The reliability of these systems are paramount when conducting research. The reliability of Myon 320 Surface Electromyography System is yet to be determined. This study aims to determine the intra-session and inter-day reliability of the Myon 320 Surface Electromyography System. Muscle activity from fifteen participants was measured at the anterior deltoid muscle during a bilateral front raise exercise, the vastus lateralis muscle
during a squat exercise and the extensor carpi radialis brevis (ECRB) muscle during an isometric handgrip task. Intra-session and inter-day reliability was calculated by intraclass correlation coefficient, standard error of measurement and coefficient of variation (CV).
The normalized root mean squared (RMS) surface electromyographic signals produced good intra-session and inter-day testing intraclass correlation coefficient values (range: 0.63–0.97) together with low standard error of measurement (range: 1.49–2.32) and CV (range: 95% Confidence Interval = 0.36–12.71) measures for the dynamic-and-isometric contractions. The findings indicate that the Myon 320 Surface Electromyography System produces good to fair reliability when examining intra-session and inter-day reliability. Findings of the study provide evidence of the reliability of electromyography between trials which is essential during clinical testing.
Original languageEnglish
Title of host publicationWearable Sensor Technology for Monitoring Training Load and Health in the Athletic Population
EditorsBilly Sperlich, Hans-Christer Holmberg, Kamiar Aminian
PublisherFrontiers Media
Pages171-177
Number of pages7
ISBN (Electronic)9782889634620
DOIs
Publication statusPublished - 29 Feb 2020

Publication series

NameFrontiers Research Topics
PublisherFrontiers
ISSN (Print)1664-8714

Keywords

  • SEMG
  • ICC
  • Squat
  • Front raise
  • Handgrip